![]() ![]() The Agilent ChemStation for LC/MSD system supports a single Agilent 1200/1100 Series LC/MSD system. I am no longer in the supervisor position, and the firm may not be willing to spend $. The Agilent ChemStation for LC/MSD systems (G2710BA) and the LC/MSD ChemStation Addon module (G2715BA) provide control, data acquisition, and data evaluation capabilities for Agilent 1200/1100 Series LC/MSD systems. I may send some old emails to you however, I am not sure if you can help. FDA wants to have the "metadata." And don't delete your data files you need to back up data and you need SOP covering that. Version B.04.02 of the Agilent ChemStation software provides LAN-based instrument control and data acquisition for LAN-capable Agilent GCs and A/D controllers. In the view of FDA, if you can print various copies of the same chromatogram from Chemstation (or Class-VP, etc), the electronic record (the chromatogram) must comply with Part 11 requirements, even if you delete the electronic data after printing. We have been requested to provide hard copies of audit trails as well. I have represented the lab to repond to FDA auditors when they asked to show the audit trails from the Dionex and Shimadzu LCs. ![]() This item may be ordered in USB or DVD format, or downloaded for a. It can not generate audit trails with Chemstation version A, unlike many versions of Chromeleon (Dionex) or Class-VP (Shimadzu). L version for MSD ChemStation and MassHunter and the Enanced NIST Version with Structures. I am really sure that the ChemStation version A is not compliant and I have used it to run the Agilent LCs. ![]()